Controller English Display Type
Keyence IM-6600E
Image Dimension Measurement System IM-6000 series
Categories: Measurement Systems, Optical Comparators
GET A PRICE
To receive a proposal, please send your request to info@sensors-trade.com and specify the required positions and company details.
Specification
Model | IM-6600E | |
Type | Controller | |
Image sensor | 1" 6.6 mega pixel CMOS | |
Display | 10.4" LCD monitor (XGA: 1024 × 768) External monitor connectable (clone output) | |
Receiver lens | Double telecentric lens | |
Image measurement | Field of view | Wide field: ø100 × 200 mm, High precision: 25 × 125 mm |
Minimum display unit | 0.1 µm | |
Repeatability | W/o stage movement | Wide field: ±1µm High precision: ±0.5µm |
With stage movement | Wide field: ±2µm High precision: ±1.5µm | |
Measurement accuracy (±2?) | W/o binding | Wide field: ±5µm*1 High precision: ±(7+0.02L)µm*2 |
With binding | Wide field: ±2µm*3 High precision: ±(4+0.02L)µm*4 | |
External remote input | No-voltage input (with contacts, without contacts) | |
External output | OK/NG/FAIL/MEAS. | Relay outputRated load 24 VDC 0.5AON resistance 50 m? or less |
Interface | LAN | RJ-45 (10BASE-T / 100BASE-TX / 1000BASE-T) |
USB 2.0 series A | 6 ports (front: 2, rear: 4) | |
Record | Hard disk drive | 250 GB |
Illumination system | Transparent | Telecentric transparent illumination (Green LED) |
Ring | Four division ring illumination (White LED) | |
External | Externally connected illumination | |
X stage | Moving range | 100 mm (electric) |
Withstand load | 2 kg | |
Z stage | Moving range | 30 mm (electric) |
Measurement points | 99 points max. (99 × 9 points possible when the function for consolidating measurement settings is used) | |
Pattern search (profi le tracking function) | XY? (with 360° rotary position compensation) | |
Pattern registration | 1000 patterns or more*5 | |
Measurement time | 2 seconds*6 | |
Basic measurement function | Distance measurement | 8 types (point-point / line-point / line-line / circle-point / circle-line / circle-circle / circle / arc) |
Angle measurement | Provided | |
Calculation | ||
Virtual line function | Point | Middle point / intersection |
Conjunction edge | Line conjunction / circular conjunction | |
Line | 6 types (midline / perpendicular line / parallel lines / tangent line / line passing through a point / approximate line) | |
Circle | Middle circle / approximate circle / auxiliary circle / inscribed and circumscribed circle | |
Application tool | Pitch measurement | Line / circumference |
Pitch angle | ||
Width measurement | Edge width | |
Thickness measurement | Thickness measurement / gap measurement between inner and outer diameters | |
Special tool | Rounded corner / curved surface / oval / reticle / point position / perimeter / area / screw / automatic measurement | |
GD&T | Shape tolerance | Straightness / circularity / profile |
Orientation tolerance | Squareness / parallelism | |
Position tolerance | Point position / concentricity | |
Element tool | Point | Point (on a line or arc) / maximum and minimum (rectangle, circle, arc) |
Line | Line / centre line / peak line | |
Circle | Circle / arc / peak circle / peak arc | |
Profi le extraction | Provided | |
Special tool | Automatic generation / gauge line | |
Manual measurement | Provided | |
Coordinate system confi guration | ||
Batch confi guration of tolerance | ||
Element list editing | ||
Measurement setting data binding function | ||
DXF export function | ||
Automatic measurement function | ||
Q-DAS data save function | ||
Measurement settings support functions | Provided*7 | |
Power supply | Power voltage | 100 to 240 VAC 50 / 60 Hz |
Power consumption | 310 mA or less | |
Environmental resistance | Ambient temperature | +10 to +35 °C |
Relative humidity | 20 to 80 % RH (No condensation) | |
Weight | Approx 8 kg | |
*1 In the range of ø80 mm from the centre of the stage within the operating temperature range of +23°C ±1°C at the focused focal point position *2 In the range of ø20 mm from the centre of the stage within the operating temperature range of +23°C ±1°C at the focused focal point position *3 In the range of ø80 × 180 mm from the centre of the stage, within the operating temperature range of +23°C ±1°C at the focused focal point position, and with a load weighing 1 kg or less on the stage (L = amount of stage movement in mm units) *4 In the range of ø20 × 120 mm from the centre of the stage, within the operating temperature range of +23°C ±1°C at the focused focal point position, and with a load weighing 1 kg or less on the stage (L = amount of stage movement in mm units) *5 Depending on the measurement setting data and number of data pieces being stored *6 Without pattern search, applied measurement, and stage movement *7 Optimum lighting guide, optimum lighting search, automatic edge extraction parameter adjustment function, and multiple edge extraction function |